Datasheet 3SP_Ethanol_1000 Package 110-202 (Spec Sensors) - 3
Hersteller | Spec Sensors |
Beschreibung | 15x15 Ethanol Sensor 1000 ppm Package 110-202 |
Seiten / Seite | 5 / 3 — 3SP_Ethanol_1000 Package 110-202. January 2017. IMPORTANT PRECAUTIONS |
Dateiformat / Größe | PDF / 503 Kb |
Dokumentensprache | Englisch |
3SP_Ethanol_1000 Package 110-202. January 2017. IMPORTANT PRECAUTIONS

Modelllinie für dieses Datenblatt
Textversion des Dokuments
3SP_Ethanol_1000 Package 110-202 January 2017 IMPORTANT PRECAUTIONS
All sensor designs are made for air monitoring @ 1 atm +/- 0.2 atm. Because applications of use and device implementation are outside our control, SPEC Sensors cannot guarantee performance in a given device or application, and disclaims any and all liability therefore.
Customers should test under their own conditions to ensure the sensors are suitable for their requirements.
Contact the factory to discuss specific concerns that might damage the sensor performance or life. • Condensation and Water (1) • Salt Water Contamination (1) • High Temperature Operation (> 70C) for more than 1 month • Low Humidity Operation (< 15% RH) for more than 3 months • High Bias voltage • Highly contaminated air over a prolonged period • High levels of particles or soot (unless proper filtering is provided) (1) Use of porous PTFE membrane or filter cap may address this concern)
MARKING INFORMATION
Sensors have serial numbers printed with individual NIST Traceable calibration data printed on each sensor. (CO version shown) PRODUCTION DATA information is current as of publication date. Products conform to specifications per the terms of the SPEC Copyright © 2011-2017, SPEC Sensors LLC Sensors standard warranty. Production processing does not necessarily include testing of all parameters.